WIZ In-situ Probe | SYSTEA
SYSTEAWIZ PROBE Request a quote
Winner · ACT Nutrient Sensor Challenge · Nitrate & Phosphate Portable in-situ probe for water analysis

Lab-grade chemistry, in the water.

WIZ brings wet-chemistry analysers into rivers, lakes, reservoirs and coastal waters. It measures up to four nutrients or other parameters on site, for months, and sends the results straight to your control centre.

4
parameters in a single probe
2.5months
unattended deployment
<1ppb
MDL for NH3-N (fluorimetric OPA)
40min
full MP4 Nutrients cycle
−10m
maximum operating depth
WIZ in action

From the buoy to the dashboard in our 45 seconds video.

See how WIZ is deployed, how the analysis runs inside the probe, and how the results reach the lab.

Independent verification

Winner of the ACT Nutrient Sensor Challenge.

Nitrate categoryPhosphate category

The Alliance for Coastal Technologies (ACT), with funding from the U.S. Integrated Ocean Observing System (IOOS), challenged sensor makers to deliver affordable, field-ready nutrient analysers. After laboratory tests and three field deployments, an independent judging panel named SYSTEA the winner in both the nitrate and phosphate categories, with the WIZ probe.

“Their instruments represent a cost-effective, commercially available solution… [and] performed well in lab testing and across the three field deployments, including 72 days in Chesapeake Bay.”

U.S. IOOS / NOAA · March 2017
Technicians lowering a WIZ probe into a monitoring buoy
A WIZ probe being installed in a monitoring buoy. The ACT award was presented at the ASLO Aquatic Sciences Meeting in Honolulu, 2017.
Fresh water

Maumee River, Ohio

Great Lakes basin, river conditions with high nutrient loads.

Brackish water

Chesapeake Bay, Maryland

Estuarine deployment from the Chesapeake Biological Laboratory pier.

Salt water

Kaneohe Bay, Hawaii

Open coastal waters with very low nutrient concentrations.

Advantages and benefits

Built for long, unattended campaigns.

WIZ combines the reliability of reference wet-chemistry methods with the autonomy of an in-situ sensor.

Multiparametric

Up to 4 parameters per unit

Ammonia, nitrate + nitrite, nitrite, phosphate, silica, TN, TP, COD/TOC, BOD, turbidity and metals, in one probe.

Flexibility

10+ ready configurations

More than ten multiparametric configurations are already developed, and methods are selected from the control software.

Low reagent use

A few µL per analysis

Microvolumes keep the reagent set small and stable for 4 to 10 weeks, depending on water temperature.

Low power

Battery operated

3 W in stand-by and 8 W on average during analysis, at 12 VDC. A deployment kit with buoy, warning lamp, batteries and solar panels is available.

Reagent canister

Swap on site

A new reagent set plugs in on site without removing WIZ from the water. Calibration runs automatically after each replacement.

Quick start-up

Delivered ready to measure

Each unit ships fully tested and calibrated to your specifications. Fill the reagent canister, connect power and start.

Autowash

Self-cleaning

WIZ washes and cleans itself at the end of each measurement. Off-scale samples are diluted and re-analysed automatically.

QC control

Quality checks in the field

Programmable QC checks while deployed, with automated sample spike and a concentrated QC solution diluted online.

Filtration

Wide filter range

10, 25, 100 or 400 µm filters, with or without self-cleaning, or a 0.25 µm self-purging microfilter.

Rendering of the WIZ quartz flow cell with fluidic channels
Patented µLFR technology

A complete analyser in a 140 mm tube.

WIZ runs sequential batch analyses on every sample with patented µLFR technology, the same chemistry used in the laboratory, scaled down to microlitres.

  • Quartz flow cell with 4 measuring sides, 10 mm × 2.5 mm
  • Sample preheating and operation from 4 to 40 °C
  • Up to 8 reagents plus QC check and wash solutions
  • NO3 reduction by UV photoreduction, with no cadmium and no hydrazine
Absorbance

Colorimetric

Multi-beam fibre-optic colorimeter with silicon detector.

Fluorimetry

Fluorimetric

LED and filters, excitation and emission at method-specific wavelengths.

Nephelometry

Turbidimetric

Scattering measurement at 800 nm.

UV 254 nm

COD UV

LED at 254 nm, corrected at 525 nm.

UV 220 / 270 nm

TN / NO3 UV

Xenon flash lamp with 220 and 270 nm filters.

Standard configurations

Choose the parameters, keep one probe.

A complete MP4 Nutrients cycle takes 40 minutes. Other configurations run in 40 to 60 minutes, at a programmable measurement frequency.

MP4 Nutrients
NH₃NO₂+NO₃NO₂PO₄
40-minute complete cycle
MP4 Nutrients & SiO₂
NH₃NO₂+NO₃PO₄SiO₂
MP5 Nutrients
NH₃NO₂+NO₃PO₄SiO₂NO₂
Requires datalogger control
MP3 Nitrogen
TNNO₂+NO₃NO₂NH₃
MP4 Phosphorus
TPP inorg.P organicOrtho-P
MP3 Phosphorus & Ammonia
TPPO₄NH₃
MP4 ECO
TOCCODBODTurbidity
Custom
Cr⁶⁺Fe²⁺Ni²⁺Cu²⁺Zn²⁺
And many others on request
MethodChemistryTypical rangeMDL
NH₃-NOPA, fluorimetric, Ex/Em 370 / 420–470 nm0–500 ppb<1
NO₃-N ColoVanadium chloride reduction to NO₂, SAA + NED0–1000 ppb<2
NO₃-N UVDirect reading at 220 nm, correction at 270 nm0–10 mg/L<0.3
NO₂-NSAA + NED0–250 ppb<1
PO₄-PMolybdate / ascorbic acid0–1000 ppb<2
SiO₂Silicomolybdate / ascorbic acid0–2000<3
TPHT digestion, UV oxidation, molybdate / ascorbic acid0–1000 ppb<3
TNUV oxidation to NO₃, UV photoreduction to NO₂, NED + SAA0–5.0 mg/L<0.05
TN UVUV oxidation to NO₃, direct reading at 220 nm0–10 mg/L<0.3
TOC / CODDirect reading at 254 nm, correction at 550 nm0–50…200 mg/L<1
BODTryptophan-like, fluorimetric, Ex/Em 280 / 360 nm0–20…500 mg/L<1
TurbidityNephelometric at 800 nm0–20…500 NTU<1

MDL in the unit of the typical range. Unique methods: NH₃ OPA fluorimetric (single reagent, high stability) and NO₃+NO₂ vanadium chloride reduction (green method, no cadmium or hydrazine).

Technicians reviewing WIZ nutrient trends on the Zetaced dashboard
Zetalog + Zetaced

Your data, live from the field.

Zetalog switches WIZ on just before each measurement cycle and off as soon as it ends, to save power. It sends the results over GPRS to Zetaced, the SYSTEA control centre on a web server.

  • Zetaced stores, validates and displays the data as tables and trend graphs
  • Remote management of the probe and of the filtration unit
  • WIZ Control Panel to set up the probe, view OD in real time, edit measuring sequences and download data over RS-232 or a GSM modem
Technical data

Specifications

Technology
µLFR (patented), sequential batch
Applications
Hydrology, water resources, coastal monitoring
Cycle time
MP4 Nutrients 40 min; other configurations 40–60 min
Measurement frequency
Programmable
Reagents
Up to 8, plus QC check and wash solutions
Reagent stability
4–10 weeks, depending on water temperature
Deployment time
Up to 2.5 months
Reading methods
Absorbance, fluorimetry, UV, nephelometry
Calibration
Automated after reagent replacement
Wastes
1 × 5 L bag for reagent wastes; washings and sample discharged freely
Body material
PVC
Max depth
−10 m
Operating temperature
4–40 °C
Analytical unit
Ø 140 × 535 mm
Reagent canister
Ø 70 × 230 mm (standard)
Power supply
12 VDC
Power consumption
3 W stand-by, 8 W average during analysis, max 1.5 A; MP4 standard 0.46 Ah
Connector
Submersible male 6-pole, RS-232 + 12 VDC
Cable
Submersible, 10 m standard
Data output
RS-232

Standard package

  • WIZ probe
  • Standard reagent canister (enlarged version on request)
  • 12 VDC power supply for maintenance and setup
  • Reagent bag filling tool
  • 10 m submersible cable
  • Carrying case
  • WIZ Control Panel software licence
  • Start-up kit

Options

  • Custom cable length
  • Enlarged reagent canister
  • Zetalog datalogger
  • Zetaced setup and licence
  • Deployment kit: buoy, warning lamp, batteries, charger, solar panels
Talk to a specialist

Plan your WIZ deployment.