Up to 4 parameters per unit
Ammonia, nitrate + nitrite, nitrite, phosphate, silica, TN, TP, COD/TOC, BOD, turbidity and metals, in one probe.
WIZ brings wet-chemistry analysers into rivers, lakes, reservoirs and coastal waters. It measures up to four nutrients or other parameters on site, for months, and sends the results straight to your control centre.
See how WIZ is deployed, how the analysis runs inside the probe, and how the results reach the lab.
The Alliance for Coastal Technologies (ACT), with funding from the U.S. Integrated Ocean Observing System (IOOS), challenged sensor makers to deliver affordable, field-ready nutrient analysers. After laboratory tests and three field deployments, an independent judging panel named SYSTEA the winner in both the nitrate and phosphate categories, with the WIZ probe.
“Their instruments represent a cost-effective, commercially available solution… [and] performed well in lab testing and across the three field deployments, including 72 days in Chesapeake Bay.”
U.S. IOOS / NOAA · March 2017

Great Lakes basin, river conditions with high nutrient loads.
Estuarine deployment from the Chesapeake Biological Laboratory pier.
Open coastal waters with very low nutrient concentrations.
WIZ combines the reliability of reference wet-chemistry methods with the autonomy of an in-situ sensor.
Ammonia, nitrate + nitrite, nitrite, phosphate, silica, TN, TP, COD/TOC, BOD, turbidity and metals, in one probe.
More than ten multiparametric configurations are already developed, and methods are selected from the control software.
Microvolumes keep the reagent set small and stable for 4 to 10 weeks, depending on water temperature.
3 W in stand-by and 8 W on average during analysis, at 12 VDC. A deployment kit with buoy, warning lamp, batteries and solar panels is available.
A new reagent set plugs in on site without removing WIZ from the water. Calibration runs automatically after each replacement.
Each unit ships fully tested and calibrated to your specifications. Fill the reagent canister, connect power and start.
WIZ washes and cleans itself at the end of each measurement. Off-scale samples are diluted and re-analysed automatically.
Programmable QC checks while deployed, with automated sample spike and a concentrated QC solution diluted online.
10, 25, 100 or 400 µm filters, with or without self-cleaning, or a 0.25 µm self-purging microfilter.

WIZ runs sequential batch analyses on every sample with patented µLFR technology, the same chemistry used in the laboratory, scaled down to microlitres.
Multi-beam fibre-optic colorimeter with silicon detector.
LED and filters, excitation and emission at method-specific wavelengths.
Scattering measurement at 800 nm.
LED at 254 nm, corrected at 525 nm.
Xenon flash lamp with 220 and 270 nm filters.
A complete MP4 Nutrients cycle takes 40 minutes. Other configurations run in 40 to 60 minutes, at a programmable measurement frequency.
| Method | Chemistry | Typical range | MDL |
|---|---|---|---|
| NH₃-N | OPA, fluorimetric, Ex/Em 370 / 420–470 nm | 0–500 ppb | <1 |
| NO₃-N Colo | Vanadium chloride reduction to NO₂, SAA + NED | 0–1000 ppb | <2 |
| NO₃-N UV | Direct reading at 220 nm, correction at 270 nm | 0–10 mg/L | <0.3 |
| NO₂-N | SAA + NED | 0–250 ppb | <1 |
| PO₄-P | Molybdate / ascorbic acid | 0–1000 ppb | <2 |
| SiO₂ | Silicomolybdate / ascorbic acid | 0–2000 | <3 |
| TP | HT digestion, UV oxidation, molybdate / ascorbic acid | 0–1000 ppb | <3 |
| TN | UV oxidation to NO₃, UV photoreduction to NO₂, NED + SAA | 0–5.0 mg/L | <0.05 |
| TN UV | UV oxidation to NO₃, direct reading at 220 nm | 0–10 mg/L | <0.3 |
| TOC / COD | Direct reading at 254 nm, correction at 550 nm | 0–50…200 mg/L | <1 |
| BOD | Tryptophan-like, fluorimetric, Ex/Em 280 / 360 nm | 0–20…500 mg/L | <1 |
| Turbidity | Nephelometric at 800 nm | 0–20…500 NTU | <1 |
MDL in the unit of the typical range. Unique methods: NH₃ OPA fluorimetric (single reagent, high stability) and NO₃+NO₂ vanadium chloride reduction (green method, no cadmium or hydrazine).

Zetalog switches WIZ on just before each measurement cycle and off as soon as it ends, to save power. It sends the results over GPRS to Zetaced, the SYSTEA control centre on a web server.
Tell us about your site and the parameters you need. A SYSTEA application specialist will reply with a proposed configuration and a quote.
info@systea.it
+39 0775 776058
SYSTEA S.p.A. · Anagni (FR), Italy