SH-FCl200 Residual chlorine digital measurement device

SH-FCl200 Residual chlorine digital measurement device

Overview

SH-FCl-200 is an on-line constant voltage digital residual chlorine measurement device. The flow path design of the product ensures constant measurement flow rate, effectively eliminating the influence of flow rate on the performance of the product; temperature and pH automatical compensation, RS485 interface and standard Modbus protocol.

 

Product features

  • With pH and temperature automatic compensation, applicable to a wide range of environments
  • Unique flow path design guarantees a constant flow rate during measurement
  • Electrodes can be replaced independently, no reagents, simple operation and low maintenance
  • Fast response time, real-time detection of residual chlorine in water
  • RS485 communication interface, standard Modbus protocol for easy integration.

 

Technical specifications

Working Principle Constant voltage glass electrode method
Measurement value Residual chlorine
pH
Temp
Measurement range Residual chlorine: (0-2/20) mg/L
0~14 pH
(0~50) °C
Accuracy Residual chlorine: ±10%; ±0.05 mg/L at concentrations below 0.5 mg/L
<0,1 pH
0.1 °C
pH comp. range 5~8.5 pH
Detection limit 0,02 mg/L
Resolution 0,01 mg/L
Response time ≤10 s
Comm. interface RS485, standard Modbus protocol
Dimensions (250*203*165) mm
Materials Electrode: gold ring glass; Detection cell: acrylic
Weight 2 Kg
Operating voltage 24 Vdc
Power absorption 0.3 W
Minimum flow rate 0,5 L/min
Cleaning method Compressed air
Calibration period 1-3 months
Operating temp. (0~50) °C

 

Product applications

Tap water, Water from waterworks and end-of-network water, secondary water supply.