SH-EC200 Conductivity digital sensor
Overview
The SH-EC-200 conductivity digital sensor has a built-in NTC temperature sensor and a compensation algorithm that automatically compensates for the effect of temperature on the measurement. The sensor uses RS485 interface and standard Modbus protocol.
Product features
- Electrode constants are stable and unaffected by polarization
- Temperature compensation function, applicable to a wide range of environments
- Additional salinity data available on request
- Internal storage of calibration data
- Corrosion-resistant housing, waterproof grade IP68, to work underwater for a long time
- RS485 communication interface, standard Modbus protocol for easy integration.
Technical specifications
Working principle | Two electrodes/four electrodes |
Measurement range |
(0~5000/100000) ms/cm, other ranges can be customized; Temperature: (0~50) °C |
Repeatability | <1 % |
Accuracy | ± 1 % |
Zero drift | ± 0.5% F.S. |
Drift (24h) | ± 1% |
Resolution | 1 us/cm |
Response time | 10 s |
Protection class | IP68 |
Comm. interface | RS485, standard Modbus protocol, baud rate 9600 |
Dimensions | (φ33*224) mm/(φ16*180) mm, cable 10m (customizable) |
Weight | 0,5 kg |
Materials | 316 Stainless-steel/PVC/POM/Titanium alloy available |
Mounting threads | NPT3/4 thread on top, NPT3/4 male thread on bottom available |
Operating voltage | 9~36 Vdc |
Power consumption | 0.4 W |
Calibration period | 2-3 months |
Max. depth | 30 meters |
Operating temp. | (0~50) °C |
Product applications
Surface water, municipal and industrial wastewater, industrial process water, drinking water